ICP-MS (Inductively Coupled Plasma Mass Spectrometry) is a highly sensitive analytical technique that can detect and quantify a variety of elements with a detection limit as low as ppt (parts per trillion). It is widely used in many fields such as environmental science, geochemistry, biomedicine, and materials science. In the ICP-MS analysis process, sample preparation is a crucial step that directly affects the accuracy and reliability of the final analysis results. Good sample preparation ensures that the sample enters the ICP-MS instrument in a suitable state, thereby achieving accurate elemental analysis.

ICP-MS DW-SUPEC7000 Inductively Coupled Plasma Mass Spectrometer Manufacturers

Why is Sample Preparation Crucial for ICP – MS?

The ICP – MS instrument itself has the capabilities of high sensitivity, wide dynamic range, and simultaneous multi – element analysis. However, the exertion of these advantages largely depends on the quality of sample preparation. Original samples often have complex compositions and physicochemical properties. For example, the organic matrix in biological samples, the complex mineral components in environmental samples, and the high – concentration matrix in industrial samples. These factors may interfere with the normal analysis process of ICP – MS, such as causing matrix effects, polyatomic ion interferences, etc.

Therefore, sample preparation is a necessary step to transform the original sample into a form suitable for ICP – MS analysis. Through appropriate sample preparation, interfering substances can be removed, sample concentration can be adjusted, and solid samples can be converted into solutions, etc., thus ensuring that ICP – MS can accurately determine the element content in the sample.

Key Steps in ICP-MS Sample Preparation

Effective sample preparation involves a sequence of carefully executed steps tailored to the sample type and the analytical goals. Each step aims to ensure the sample is compatible with the ICP-MS system and free of interferences. Here are the key aspects:

icp ms sample preparation

1. Preparation of Tools and Equipment for Sample Preparation

The tools and equipment used during sample preparation must be meticulously cleaned and prepared to minimize contamination and ensure reliable results.

Clean and Decontaminate Equipment:

Use high-purity acids (e.g., nitric acid) to clean labware such as digestion vessels, centrifuge tubes, and pipettes.

Rinse all tools thoroughly with ultra-pure water to remove any acid residues or particulates.

Dry labware in a clean environment, such as a laminar flow hood, to avoid contamination from airborne particles.

Inspect Tools:

Check equipment for cracks, scratches, or residues that could trap contaminants.

Replace disposable items, such as pipette tips or filters, after each use to maintain purity.

Calibrate Instruments:

Ensure all equipment, such as microwave digesters, centrifuges, and balances, is calibrated according to manufacturer specifications.

Test filtration systems and nebulizers for clogs or inefficiencies before starting the preparation process.

Preparation of tools and equipment is a critical step that lays the groundwork for accurate and consistent ICP-MS analyses. Proper cleaning, inspection, and calibration help eliminate contamination risks and maintain system efficiency.

Microwave Digestion System

2. Sample Collection and Storage

The integrity of a sample is directly influenced by how it is collected and stored. Using proper techniques ensures the sample remains uncontaminated and stable until analysis.

The methods and tools used to collect samples depend on the type of sample. For example, for environmental water samples, clean, uncontaminated sampling bottles are required to be collected at specific locations and depths, and should collect in a single motion to avoid stratification and contamination from multiple dips; for biological tissue samples, sterile instruments are required for collection. The amount of sample collected also needs to be determined based on the needs and detection limits of subsequent analysis.

In terms of sample preservation, the stability of the sample needs to be considered. For example, some water samples may require the addition of acid to inhibit microbial growth and prevent element precipitation; biological samples may need to be stored at low temperatures or with specific preservatives.

Labeling the sample is also important. This need clearly label containers with sample IDs, collection dates, and any other relevant information. At the same time, maintain detailed records of collection methods and storage conditions for traceability.

Proper sample collection and storage is the first step to ensure accurate analytical results, which can prevent the sample from deteriorating or losing elements before it arrives at the laboratory for preparation and analysis.

ICP-MS2000 (4)

3. Sample Pretreatment

Sample pretreatment involves preparing the raw sample into a form compatible with ICP-MS analysis. This step is crucial for ensuring that the sample is free of particulates, dissolved completely, and within the detection range of the instrument. Key methods including:

Dissolution: For solid samples, such as soil, rock or metal materials, they need to be dissolved into a solution for ICP-MS analysis. Common dissolution methods include acid dissolution (such as nitric acid, hydrochloric acid, hydrofluoric acid, etc.) and alkaline dissolution. Acid dissolution can select a single acid or a mixed acid according to the nature of the sample. For example, nitric acid-hydrochloric acid mixed acid (aqua regia) can be used to dissolve precious metals. During the dissolution process, the reaction conditions, such as temperature, time and acid concentration, need to be controlled.

Digestion: For samples containing organic matter, such as biological tissues or foods, digestion is a common pretreatment method. The purpose of digestion is to destroy organic matter and release the elements in it. Common digestion methods include dry ash method, wet digestion and microwave digestion.

  • Dry ash method is to ash the sample at high temperature and then dissolve the ash with acid;
  • Wet digestion is to use acid solution under heating conditions to oxidize and decompose the organic matter in the sample;
  • Microwave digestion is to use microwave radiation to accelerate the digestion reaction, which has the advantages of fast, high efficiency and low reagent usage.

Separation and enrichment: When the target element content in the sample is low or there is severe interference, separation and enrichment are required. For example, ion exchange resin can be used to selectively adsorb the target ions, thereby achieving separation from interfering ions; co-precipitation can precipitate the target element together with the precipitant to achieve the purpose of enrichment.

Dilution: Dilute samples to bring element concentrations within the instrument’s dynamic range. Use ultra-pure water or matrix-matched diluents to maintain consistency and minimize matrix effects.

Matrix Matching: Prepare calibration standards that closely match the matrix composition of the samples to compensate for potential matrix effects. This step minimizes interferences from the matrix and ensures more accurate quantification.

Sample pretreatment is an important step in converting the original sample into a solution suitable for ICP-MS analysis. Different sample types and analysis purposes require the selection of appropriate pretreatment methods.

Common Challenges in ICP-MS Sample Preparation and How to Overcome Them

Despite careful planning, sample preparation for ICP-MS can pose several challenges. Recognizing these obstacles and addressing them proactively ensures high-quality analyses.

1. Matrix effect

Matrix effect refers to the influence of other components in the sample other than the target element on the target element analysis signal. For example, in a high-salt sample, a large amount of salt matrix may inhibit the ionization efficiency of the target element, resulting in a decrease in the signal; or generate polyatomic ion interference, making the analysis result inaccurate.

Methods to overcome:

  • Matrix matching: Prepare a standard solution similar to the sample matrix so that the standard solution and the sample are affected by the same matrix effect during the analysis process, thereby improving the accuracy of the analysis result.
  • Internal standard method: Select a suitable internal standard element and add the same amount of internal standard element to the sample and standard solution. The internal standard element should have similar physical and chemical properties to the target element. During the analysis process, the internal standard element and the target element are subject to the same interference. The signal of the target element is corrected by the change of the internal standard element signal, thereby eliminating the influence of the matrix effect.
ICP-MS Sample Pretreatment

2. Polyatomic ion interference

In ICP-MS analysis, atoms, ions and molecules in the plasma may react to form polyatomic ions. These polyatomic ions may have the same mass-to-charge ratio as the target element, thereby interfering with the detection of the target element. For example, argon (Ar) is a commonly used plasma gas in ICP-MS, which may react with elements or other gases in the sample to generate polyatomic ions, such as ArO +, ArCl +, etc.

Overcoming methods:

  • Optimize instrument parameters: By adjusting the parameters of the ICP-MS instrument such as the RF power, nebulizer gas flow rate, sampling depth, etc., the properties of the plasma can be changed to reduce the formation of polyatomic ions. For example, appropriately reducing the RF power can reduce the generation of polyatomic ions, but it may affect the ionization efficiency of the element, and a balance needs to be found between the two.
  • Select the right isotope: For elements with polyatomic ion interference, isotopes with less interference can be selected for analysis. For example, for the element iron (Fe), 56Fe may be interfered by polyatomic ions, while the interference of 57Fe is relatively small, so 57Fe can be selected for analysis.

3. Contamination

Contamination from labware, reagents, and the environment can significantly impact results. Use high-purity reagents, clean laboratory practices, and proper sample storage to minimize contamination.

4. Incomplete Digestion

Ensure complete dissolution of samples by optimizing digestion conditions and using appropriate acid mixtures.

5. Memory Effects

These occur when residual analyte from previous samples affects subsequent analyses. Thorough cleaning of the sample introduction system and use of appropriate cleaning solutions can help prevent memory effects.

Tips to ensure accuracy and consistency

Consistency is key to producing reliable ICP-MS results. The following tips help maintain uniformity across sample preparation batches.

  • Use of quality control samples: During the sample preparation process, regularly analyze quality control samples such as standard reference materials (SRMs) and spiked samples. Standard reference materials have known elemental content and can be used to verify the accuracy of sample preparation and analytical methods; spiked samples are samples with known amounts of target elements added to them, and the accuracy of the method is evaluated by analyzing the recovery of spiked samples.
  • Strict operating specifications: Implement and document standard operating procedures (SOPs) for all preparation steps, including the use of reagents, instrument operation, sample processing steps, etc. For example, when using a pipette to transfer reagents, ensure the accuracy of the pipette to avoid errors; when digesting samples, strictly control the digestion conditions, such as temperature and time.
  • Data recording and review: Record all data in the sample preparation process in detail, including sample collection information, pretreatment methods, reagents used, and instrument parameters. Review the recorded data to promptly detect and correct possible errors.

Sample preparation is an indispensable and important part of ICP-MS analysis. From sample collection and preservation, to pretreatment, each step has a key impact on the final analysis results. Proper sample preparation can give full play to the high sensitivity and multi-element analysis advantages of ICP-MS, and provide reliable elemental analysis data for scientific research, environmental monitoring, quality control, etc. in various fields.

Related Products Recommendation