DW-GR Series Vertical Autoclave
- DW-GR-DA: Fully automatic with fast cooling fan
- DW-GR-DF: DW-GR-DA+ automatic water feeding
- DW-GR-DR: DW-GR-DF+fully automatic drying function
Advanced Functions of DW-GR Series Vertical Autoclave
- Smart microcomputer control
“Inspiration” high speed microcomputer control, not only to monitor and control the whole process, but also to update the function in the future.
- Program Memory storage system
User can set their own program, and save it in the system.
- Calibration interface
Offer temp. and pressure interface, together with the 3Q adapter joints, can insert max 15 sensors one time to do the calibration work
- Saturated steam monitoring
The system automatically monitors the purge out of cold air to ensure the pure steam environmentk
- Auto water fill(only for DF, DR models)
Equipped with automatic water feeding function, standard is with 28L water tank at the back to store water, the option is direct water supply by pipe connection from water source while manufacturing.
- Biosafety function(only for DAS, DFS, DRS)
Use a 0.2 μm PTFE filter packed in housing to ensure that no biohazard aerosols, microorganisms, and viruses are passing out through the filter during exhausting, so as to make sure the biology DAfety of the lab.
- Monitor of exhaust filter( only for DAS, DFS, DRS)
Exhaust filter is put inside a pressure-resistant housing (stainless steel material) with an extra PT100 temperature sensor to monitor the temperature, so as to make sure the filter and housing are also being sterilized thoroughly, and will not lead to secondary biohazard pollution.
- SUS316L Stainless steel(Option)
Stronger corrosion, longer service life, and greater pressure resistance.
- Humanized permission management
The management of the autoclave is classified into 3 levels,user-administrator- engineer, to make sure proper proper usage.
- Agar program
Can heat up, melt, warm the agar, or be used as a chamber preheating function
- Waste program
Special waste program to sterilize the waste thoroughly
- Auto start-up appointment
Can make an appointment in 0- 15 days
- Fully automatic drying( only for DR models)
Can set drying temp. and drying time. After sterilization, water will be drained out automatically, then an inbuilt pump will supply clean air into the chamber through an air filter.
- Air filter( only for DR models)
Filtration is higher than 99.99% in particles with a diameter of 0.2 μm.
- LCD/USB/APP/Smart management system(Option)
- LCD:5 inch LCD screen to show information in words
- USB: to help to store and transfer the sterilization data
- APP: Remote management APP(only in China)
- Smart management system: linked with LIMIS/SCADA system(only in China)
- Auto drainage(option)/Auto cleaning(only for DF/DR)
Added with a water pump, press the drainage button to start drainage, or set it to timed automatic drainage.
- Condenser(Option)
Equipped with a professional condenser, the steam generated during the sterilization process is converted into water through the condenser, achieving zero discharge of water vapor/mist.
Safety Features of DW-GR Series Vertical Autoclave
- Humanized design: The lid and bench corners are covered with insulation to avoid getting scalded.
- Interlocking device: The exclusive patented “eight-column evenly distributed” interlock device is safer and more reliable than ordinary interlocking devices.
- Lid closing checking: The Machine can not start if the lid is not locking well by checking mechanically and electrically.
- Real-time monitor: The system will monitor the lid locking, pressure, and temperature status on a real-time basis to guarantee the safe and successful work of the autoclave.
- Cooling lock OPEN temperature: Each program can set different lid OPEN temperatures, especially for the liquid program, so as to prevent the user from opening the lid and causing burns when the articles are still in a high-temperature state
- Triple protection for lack of water: The bottom of the sterilization chamber is equipped with three different dry-burning protection devices: liquid expansion type, copper temperature sensing type, and electric ion type (water level sensor), to avoid misjudgment caused by a single method.
- Double protection for overpressure: Mechanical pressure gauge and electric sensing system, once pressure is abnormal, the system will cut power, release the pressure, and send out an alarm.
- Auto troubleshooting: The system will monitor the working status, and will cut the power, send out an alarm, and report the error information once any abnormal situation is detected safety test function: Regular testing of system security and safety protection measures is possible.
- Electrical protection: Equipped with a highly sensitive electrical protection system, effectively providing leakage protection, overcurrent and short-circuit protection.
Specification Table of DW-GR Series Vertical Autoclave
Model | Standard DW-GR60DA | Standard DW-GR85DA | Standard DW-GR110DA |
Auto water fill DW-GR60DF | Auto water fill DW-GR85DF | Auto water fill DW-GR110DF | |
Auto drying DW-GR60DR | Auto drying DW-GR85DR | Auto drying DW-GR110DR | |
Difference | DW-GR-DA: Fully automatic with fast cooling fan DW-GR-DF: DW-GR-DA+ automatic water feeding DW-GR-DR: DW-GR-DF+fully automatic drying function | ||
Capacity(L) | 60 | 85 | 110 |
Dimension(L*W*H)(mm) | DA:644x660x980 | DA: 644x660x980 | DA:644x660x980 |
DF/DR:644x831x980 | DF/DR:644x831x980 | DF/DR: 644x831x1180 | |
Chamber dimension (Dia*H)(mm) | Φ390×505 | Φ390×700 | Φ390×895 |
Rated power(W) | 2900 | 4600(2900 as option) | 4600 |
SUS 304 Basket(Dia*H)(mm)(Note 1) | (Φ370×190)x2 | (Φ370×280)x2 | (Φ370×250)x3 |
Power requirement | 220V±10% 16A 50HZ/60HZ | 220V±10% 32A 50HZ/60HZ | 220V±10% 32A 50HZ/60HZ |
Temperature display accuracy | 0.1℃ | ||
Working environment | 5℃~40℃ relative humidity ≤85% | ||
Chamber material | SUS304 stainless steel | ||
Sterilization temp.range | 105℃~138℃ | ||
Sterilization time range | 1~6000 min | ||
Melting temp. range | 60℃~115℃ | ||
Melting time | 1-6000 min | ||
Warming temp | 45℃~79℃ | ||
Warming time | 1~9999 min | ||
Drying(only for DR) | Drying time:1~300 min Drying temp:80~160℃ | ||
Cooling lock OPEN temp.range | 40℃~99℃ for solid, waste, agar, self-defined 40℃~80℃(note 2)for liquid | ||
Sterilization mode | Liquid, Liquid with warming, solid, Wrapped instrument, Fabric, Rubber, Fast, Waste, Agar, Self-defined | ||
Program number saved | DA: 20 programs DF/DR: 60 programs | ||
Exhaust speed control | Fully automatic internal discharging with 6 levels choice | ||
Lid open way | Flip open(up and down) | ||
Water tank | DA:2 water tanks to collect steamDF:l water storage tank more than DADR:1 water tank to collect steam+1 water storage tank (note 3) | ||
Appointment timer range | (Auto start up appointment)0~15 days | ||
Pressure gauge range | 0.1~0.5MPa | ||
Pressure valve set pressure | 0.29 Mpa | ||
Controller | “Inspiration” smart microcomputer controller | ||
Safety Device | Eight column evenly distributed “interlock device electric double inter bock lid closing checking dry scorch protection. age), cooling lock, anti-scald lid and bench corner, automatic troubleshooting | ||
Biosafety function | Option for 0.2um biosafety filter(basic model+S) |
- Note 1:Baskets can also be customized
- Note 2:Cooling lock OPEN temp. is related to local boiling point
- Note3:Optional sewer drainage for DR while manufacturing